Human CMV IgM
FOR RESEARCH USE ONLY
96 determinations
Purpose
This kit allows for the determination of CMV IgM expression in Human serum, and other biological fluids.
Principle of the assay
The kit assay CMV IgM level in the sample, use Purified antigen to coat microtiter plate wells, make solid-phase antigen, then add CMV IgM to wells, Combined With CMV IgM, after washing and removing non-combinative antibody and other components ,then Combined antigen which with HRP labeled become antigen–antibody-enzyme-antigen complex, after washing Compley, Add TMB substrate solution,, TMB substrate becomes blue color At HRP enzyme-catalyzed, reaction is terminated by the addition of a sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450 nm. Compared with the CUTOFF value, according to this to judge CMV IgM exist in the sample or not.
Materials provided with the kit
1
wash solution
20ml×1bottle
7
Stop Solution
6ml×1 bottle
2
HRP-Conjugate reagent
6ml×1 bottle
8
Positive control
0.5ml×1 bottle
3
Microelisa stripplate
12well×8strips
9
Negative control
0.5ml×1bottle
4
Sample diluent
6ml×1 bottle
10
Instruction
1
5
Chromogen Solution A
6ml×1 bottle
11
Closure plate membrane
2
6
Chromogen Solution B
6ml×1 bottle
12
Sealed bags
1
Specimen requirements
- extract as soon as possible after Specimen collection,and according to the relevant literature, and should be experiment as soon as possible after the extraction. If it can’t, specimen can be kept in -20 ℃ to preserve, Avoid repeated freeze-thaw cycles.
- Can’t detect the sample which contain NaN3, because NaN3 inhibits HRP active.
Assay procedure
1.Number: to sample correspond microtitration well and Number Sequence, each plate should be set feminine comparison 2 wells, masculine comparison 2 wells, blank comparison 1 well(don’t add sample and HRP-Conjugate reagent to blank comparison well, other each step the operation are same).
2.add sample:separay add Positive control and Negative control 50μl to the Positive and Negative well . add Sample dilution 40μl to testing sample well, then add testing sample 10μl. add sample to the bottom of ELISA plates coated well , don’t touch the well wall as far as possible, and Gently mix.
3.Incubate: After closing plate with Closure plate membrane ,incubate for 30 min at 37℃.
4.Configurate liquid: 30-fold (or 20-fold)wash solution diluted 30-fold (or 20-fold) with distilled water until 600ml,and reserve.
5.washing:Uncover Closure plate membrane, discard Liquid, dry by swing, add washing buffer to every well, still for 30s then drain, repeat 5 times, dry by pat.
6.add enzyme:Add HRP-Conjugate reagent 50μlto each well, except the blank well.
7.incubate:Operation with 3.
8.washing:Operation with 5.
9.color:Add Chromogen Solution A 50ul and Chromogen Solution B to each well, evade the light preservation for 15 min at 37℃
10.Stop the reaction:Add Stop Solution50μl to each well, Stop the reaction(the blue color change to yellow color).
11. assay:take blank well as zero , Read absorbance at 450nm after Adding Stop Solution and within 15min.
Determine the result
Test validity: the average of Positive control well≥1.00; the average of Negative control well ≤0.10.
Calculate Critical(CUT OFF) : Critical= the average of Negative control well + 0.15.
Negative control: sample OD< Calculate Critical(CUT OFF) is CMV IgM Negative control.
Positive control: ample OD≥ Calculate Critical(CUT OFF) is CMV IgM Positive control.
Important notes
1.Please according to use instruction strictly, Do not mix reagents with those from other lots.
2.The kit takes out from the refrigeration environment should be balanced 15-30 minutes in the room temperature then use, ELISA plates coated if has not use up after opened, the plate should be stored in Sealed bag.
3.washing buffer will Crystallization separation, it can be heated the water helps dissolve when dilute . Washing does not affect the result.
4.Closure plate membrane only limits the disposable use, in order to avoid the overlapping pollution
5.The substrate please evade the light preservation.
6.The test result determination must take the microtiter plate reader as a standard, when use dual-wavelength to assay, Reference wavelength is 630nm.
7.All samples, washing buffer and each kind of reject should according to infective material process. Stopp Solution is 2M sulphuric acid. You must pay attention to safe when use .
Storage and validity
1.Storage: 2-8℃.
2.validity: six months.
- 凡本网注明"来源:环保在线"的所有作品,版权均属于环保在线,转载请必须注明环保在线,https://www.hbzhan.com。违反者本网将追究相关法律责任。
- 企业发布的公司新闻、技术文章、资料下载等内容,如涉及侵权、违规遭投诉的,一律由发布企业自行承担责任,本网有权删除内容并追溯责任。
- 本网转载并注明自其它来源的作品,目的在于传递更多信息,并不代表本网赞同其观点或证实其内容的真实性,不承担此类作品侵权行为的直接责任及连带责任。其他媒体、网站或个人从本网转载时,必须保留本网注明的作品来源,并自负版权等法律责任。
- 如涉及作品内容、版权等问题,请在作品发表之日起一周内与本网联系,否则视为放弃相关权利。