齐一生物科技(上海)有限公司作者
Human (HMGB-1)试剂盒
资料类型 | png文件 | 资料大小 | 23828 |
下载次数 | 43 | 资料图片 | 【点击查看】 |
上 传 人 | 齐一生物科技(上海)有限公司 | 需要积分 | 0 |
关 键 词 | Human (HMGB-1)试剂盒 |
- 【资料简介】
1
Human high mobility group protein B1
( (( (HMGB-1) )) )
FOR RESEARCH USE ONLY
Assay range: :: :8µg/L -160µg/L 96 determinations
Purpose
This kit allows for the determination of HMGB-1 concentrations in Human serum,
cell culture supernates and other biological fluids
Principle of the assay
The kit assay Human HMGB-1 level in the sample, use Purified Human HMGB-1
antibody to coat microtiter plate wells, make solid-phase antibody, then add HMGB-1 to wells,
Combined HMGB-1 antibody which With HRP labeled, become antibody - antigen -
enzyme-antibody complex, after washing Compley, Add TMB substrate solution,TMB
substrate becomes blue color At HRP enzyme-catalyzed, reaction is terminated by the addition
of a sulphuric acid solution and the color change is measured spectrophotometrically at a
wavelength of 450 nm. The concentration of Human HMGB-1 in the samples is then
determined by comparing the O.D. of the samples to the standard curve.
Materials provided with the kit
1 wash solution 20ml×1bottle 7 Stop Solution 6ml×1 bottle
2 HRP-Conjugate reagent 6ml×1 bottle 8 Standard (320µg/L) 0.5ml×1 bottle
3 Microelisa stripplate 12well×8strips 9 Standard diluent 1.5ml×1bottle
4 Sample diluent 6ml×1 bottle 10 Instruction 1
5 Chromogen Solution A 6ml×1 bottle 11
Closure plate
membrane
2
6 Chromogen Solution B 6ml×1 bottle 12 Sealed bags 1
Specimen requirements 2
1. extract as soon as possible after Specimen collection,and according to the relevant
literature, and should be experiment as soon as possible after the extraction. If it can’t,
specimen can be kept in -20 ℃ to preserve, Avoid repeated freeze-thaw cycles.
2. Can’t detect the sample which contain NaN3, because NaN3 inhibits HRP active.
Assay procedure
1. Dilute and add sample:Dilute Original density Standard as follow table:
160µg/L
5 Standard 150µl Original density Standard+150µl Standard diluent
80µg/L
4 Standard 150µl 5 Standard+150µl Standard diluent
40µg/L
3 Standard 150µl 4 Standard+150µl Standard diluent
20µg/L
2 Standard 150µl 3 Standard +150µl Standard diluent
10µg/L
1 Standard 150µl 2 Standard +150µl Standard diluent
2. Add sample: Set blank wells separay (blank comparison wells don’t add sample and
HRP-Conjugate reagent, other each step operation is same). testing sample well. add Sample
dilution 40µl to testing sample well, then add testing sample 10µl (sample final dilution is
5-fold), add sample to wells , don’t touch the well wall as far as possible, and Gently mix.
3. Incubate: After closing plate with Closure plate membrane , incubate for 30 min at 37℃.
4. Configurate liquid: 30-fold(or 20-fold) wash solution diluted 30-fold (or 20-fold) with distilled
water and reserve.
5. Washing: Uncover Closure plate membrane, discard Liquid, dry by swing, add washing
buffer to every well, still for 30s then drain, repeat 5 times, dry by pat.
6. Add enzyme: Add HRP-Conjugate reagent 50µl to each well, except blank well.
7. Incubate: Operation with 3.
8. Washing: Operation with 5.
9. Color: Add Chromogen Solution A 50ul and Chromogen Solution B to each well, evade the
light preservation for 15 min at 37℃
10. Stop the reaction: Add Stop Solution50µl to each well, Stop the reaction(the blue color
change to yellow color).
11. Assay: take blank well as zero , Read absorbance at 450nm after Adding Stop Solution and
within 15min. 3
Steps description
Standard, Sample diluent
Add Standard, Sample diluent, incubate for 30 min at 37℃.
Wash 5 time,Add HRP-Conjugate reagent, incubate for 30 min at 37℃.
Wash 5 times,Add Chromogen Solution A and B, incubate for 30 min at 37℃.
Add Stopp Solution
Read absorbance at 450nm within 15 min
calculate
Calculate
Take the standard density as the horizontal, the OD value for the vertical ,draw the
standard curve on graph paper, Find out the corresponding density according to the sample
OD value by the Sample curve, multiplied by the dilution multiple, or calculate the straight line
regression equation of the standard curve with the standard density and the OD value ,with the 4
sample OD value in the equation, calculate the sample density, multiplied by the dilution factor,
the result is the sample actual density.
Important notes
1. The kit takes out from the refrigeration environment should be balanced 15-30 minutes in
the room temperature, ELISA plates coated if has not use up after opened, the plate should
be stored in Sealed bag.
2. washing buffer will Crystallization separation, it can be heated the water helps dissolve
when dilute . Washing does not affect the result.
3. add Sample with sampler Each step, And proofread its accuracy frequently, avoids the
experimental error. add sample within 5 min, if the number of sample is much , recommend
to use Volley .
4. if the testing material content is excessively higher (The sample OD is bigger than the first
standard well ),please dilute Sample (n-fold), Please diluente and multiplied by the dilution
factor.(×n×5).
5. Closure plate membrane only limits the disposable use, to avoid cross-contamination.
6. The substrate evade the light preservation.
7. Please according to use instruction strictly, The test result determination must take the
microtiter plate reader as a standard.
8. All samples, washing buffer and each kind of reject should according to infective material
process.
9. Do not mix reagents with those from other lots.
Storage and validity
1.Storage: 2-8℃.
2.validity: six months
上一篇:保温管,型能实验
下一篇:85026-55-7,络缌对照品
- 凡本网注明"来源:环保在线"的所有作品,版权均属于环保在线,转载请必须注明环保在线,https://www.hbzhan.com。违反者本网将追究相关法律责任。
- 企业发布的公司新闻、技术文章、资料下载等内容,如涉及侵权、违规遭投诉的,一律由发布企业自行承担责任,本网有权删除内容并追溯责任。
- 本网转载并注明自其它来源的作品,目的在于传递更多信息,并不代表本网赞同其观点或证实其内容的真实性,不承担此类作品侵权行为的直接责任及连带责任。其他媒体、网站或个人从本网转载时,必须保留本网注明的作品来源,并自负版权等法律责任。
- 如涉及作品内容、版权等问题,请在作品发表之日起一周内与本网联系,否则视为放弃相关权利。